If you get a Verify SCXI Chassis message box showing the SCXI
chassis model number, Chassis ID: x, and the message Slot
Number: x Configuration has module: SCXI-XXXX or
1141/1142/1143, hardware in chassis is: SCXI-YYYY,
1141/1142/1143, or Empty, complete the following troubleshooting
steps to correct the error.
1. Expand NI-DAQmx Devices.
2. Right-click the SCXI chassis and click Properties to load the
chassis configurator.
3. Under the Modules tab, ensure that the cabled module is listed in
the correct slot.
4. If the cabled module is not listed in the correct slot, complete the
following troubleshooting steps:
a. If the cabled module is not listed in the correct slot and the
slot is empty, click the drop-down listbox next to the correct
slot and select the cabled module. Configure the cabled
module following the steps listed in the SCXI Quick Start
Guide. Click OK.
b. If another module is displayed where the cabled module
should be, click the drop-down listbox next to the correct slot
and select the cabled module. A message box opens asking
you to confirm the module replacement. Click OK. Configure
the cabled module following the steps listed in the SCXI
Quick Start Guide. Click OK.
If you have more than one kind of SCXI module in the SCXI chassis,
ensure that you have the highest priority SCXI module cabled to the
E/M Series DAQ device. Refer to the SCXI Quick Start Guide to find
out which SCXI module in the chassis should be cabled to the
E/M Series DAQ device.
After checking the preceding items, return to the Troubleshooting the
Self-Test Verification section and retest the SCXI chassis.
If these measures do not successfully configure the SCXI system, contact
NI. Refer to the Technical Support Information document for contact
information.