Omega RD100B Microcassette Recorder User Manual


 
7-6
7.4 Setting the RJC Function on Channels Set to
TC Input
This section explains the details of setting RJC on channels set to TC input.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1. Hold down the
MENU
key for 3 seconds to enter Setting mode.
2. Hold down the and keys simultaneously for 3 seconds to enter Basic
Setting mode.
3. Press the
key or SHIFT + key to select RJC and then press the
key.
4. Set each item and press the
key.
Use the key or SHIFT + key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5. When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Set the channel range.
Select the RJC mode of the
thermocouple input.
Set the compensation voltage when
using the external RJC function.
The new setting takes effect.
ESC/?
First channel
Last channel
(When RJC is set to External)
01-01 Channel
Setting complete
Basic=RJC
CH=01-01
RJC=Internal
Volt= 0
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT + key to select End and then press the key. Press
the key to select Store and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.