9.4 Setting the Constants Used in Equations
This section explains the details of setting constants to be used in computing equations.
Procedure
1. Hold down the
MENU
key for 3 seconds to enter Setting mode.
2. Press the
key or SHIFT + key to select Math and then press the
key.
3. Press the key or SHIFT + key to select Constant and then press the
key.
4. Set each item and press the key.
Use the key or SHIFT + key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
5. When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
6. Hold down the
MENU
key for 3 seconds to return to Operation mode.
Select the constant number.
The new setting takes effect.
ESC/?
Set the value.
K01
Setting complete
Set=Math
Math=Constant
No.=K01
K01=1
Explanation
Constant
Up to 30 constants can be registered. Enter the values as K01 to K30 in the equations.
Set the constants in the range shown below. The maximum number of significant digits
is 5.
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29