Princeton 4300 Projector User Manual


 
Chapter 4 System Setup 37
37
Figure 12. RSConfig dialog
4. Open the WinX application and, from Setup|Hardware…, run the Hardware
Setup wizard.
5. When the PVCAM dialog (Figure 13) is displayed, click in the Yes radio button,
click on Next and continue through the wizard. After the wizard is finished, the
Controller/Camera tab card will be displayed with the Use PVCAM checkbox
selected. You should now be able to set up experiments and acquire data.
Figure 13. Hardware Setup wizard: PVCAM dialog
6. Run the software in focus mode to verify communication between the ST-133
and the host computer.
Making the Circulator-Camera Connections
1. Set up the coolant circulator according to the directions in the user manual for that
equipment. Do not apply power to the circulator until directed to do so.
2. Make the hose connections between the circulator and the camera. For best cooling
performance, the tubing should be no longer than necessary.
PI-SCX:1300 Use 1/4", thin-wall plastic tubing. Be sure the tubing is properly
secured at both ends. Note that the ports on this camera use a ferrule-less quick-
disconnect method of securing the tubing and that both the camera's valve body
and the fitting insert include automatic shutoff to prevent coolant leaks when
disconnected. For best cooling, connect the inflow and outflow tubing to the
ports as indicated in Figure 14.