Xantrex Technology XDC 10-1200 Power Supply User Manual


 
Operation
3–46 TM-XDOP-01XN
To set the voltage limits:
Step # Do This You Will See
1 ERROR MSGS
2
until you see the “SETPT LIMIT”
option
SETPT LIMIT
3
Voltage Limit
4 High 0V
5 Use the Voltage knob, arrow keys, or
numeric keypad to enter a value. The value
must be within 0V to 103% of the unit’s
rated voltage, and must be higher than the
voltage setpoint.
High #####V
6 Low 0V
The Maximum setting is
saved and the Minimum
setting appears.
7 Use the Voltage knob, arrow keys, or
numeric keypad to enter a value. The value
must be within 0V to 103% of the unit’s
rated voltage and must be lower than the
voltage setpoint.
Low #####V
8 The setting is saved and
the display returns to its
default operating mode.
MENU
ENTER
ENTER
ENTER
ENTER
TM-XDOP-01XN.book Page 46 Monday, July 17, 2006 11:19 AM