Xantrex Technology XDC 10-1200 Power Supply User Manual


 
Remote Operation
4–52 TM-XDOP-01XN
Table 4-15
QUEStionable Status Register
Bit Bit Weight Bit Name Description
0 1 VOLTage Summary Reflects a summary of the VOLTage Sub-Register.
1 2 CURRent Summary Reflects a summary of the CURRent Sub-Register.
2 4 TIME Not implemented
3 8 POWer Summary Reflects a summary of the POWer Sub-Register.
4 16 TEMPerature Summary Reflects a summary of the TEMPerature Sub-Register.
5 32 FREQuency Summary Not implemented
6 64 PHASe Summary Not implemented
7 128 MODulation Summary Not implemented
8 256 CALibration Indicates an error in the unit calibration.
9 512 Not Used Not implemented
10 1024 Not Used Not implemented
11 2048 AC Off Indicates an AC Supply failure.
12 4096 UNRegulated Indicates that the output is not regulated in either Constant
Voltage mode, Constant Current mode or Constant Power
mode. Reflects the inverse of the Operation Regulation
Summary bit.
13 8192 INSTrument Summary Not implemented
14 16384 Command Warning Not implemented
15 32768 Not Used Always zero
TM-XDOP-01XN.book Page 52 Monday, July 17, 2006 11:19 AM