No. 99MBA038A
5-4
No. Operation Screen display and remarks
8
Completion of circle (B) measurement
After the desired measured points have been entered,
press the [F5] function key corresponding to
.
–
9
Results display
The measurement results for circle (B) are displayed.
N0002
X = 6.083
Y = 6.012
D = 4.000
10
Measuring angle (C)
Measure four points on the two lines forming angle
(C).
(C)
(1)
(2)
(4)
(3)
<Remarks>
The icon corresponding to the automatically
recognized basic feature type is highlighted.
11
Completion of angle (C) measurement
After the desired measured points have been entered,
press the [F5] function key corresponding to
.
–
12
Results display
The measurement results for angle (C) are displayed.
N0003
X = 0.083
Y = 0.012
A1 = 119.4837
A2 = 240.1123
TIP
If the automatically recognized feature type is not appropriate, the feature type can be
manually corrected. See Chapter 4 of the “Software Guide” for information on how to
correct feature type.