Agilent Technologies E1364A Switch User Manual


 
ARM
The ARM subsystem selects the number of scanning cycles (1 to 32767) for
each INITiate command.
Subsystem Syntax ARM
:COUNt <number> MIN | MAX
:COUNt? [MIN | MAX]
:COUNt ARM:COUNt <number> MIN | MAX allows scanning cycles to occur a
multiple of times (1 to 32767) with one INITiate command when
INITiate:CONTinuous OFF | 0 is set. MIN sets 1 cycle, MAX sets 32767
cycles.
Parameters
Parameter Name Parameter Type Range of Values
<number> numeric 1-32767 l MIN l MAX
Comments
Number of Scans: Use only values between 1 and 32767 for the
number of scanning cycles.
Related Commands: ABORt, INITiate[:IMMediate]
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles:
This example sets a Form C Switch for 10 scans of channels 00 through 03.
When the scan sequence completes, channels 00 through 03 (relays 00
through 03) are closed.
ARM COUN 10 Set 10 scans per INIT command
SCAN (@100:103) Scan channels 00-03
INIT Start scan, close channel 00
Chapter 5 Agilent E1364A Form C Switch Command Reference 45