Agilent Technologies E1364A Switch User Manual


 
CLOSe? [ROUTe:] CLOSe? <channel_list> returns the current state of the
channel(s) queried. channel_list has the form (@ccnn) (see
[ROUTe:]CLOSe for definition). The command returns 1 if channel(s) are
closed or returns 0 if channel(s) are open.
Comments
Query is Software Readback: The [ROUTe:]CLOSe? command
returns the current software state of the channel(s) specified. It does
not account for relay hardware failures. A maximum of 127 channels
at a time can be queried for a multi-module switchbox.
Example Query Channel Closure:
This example closes channels 100 and 213 of a two-module switchbox and
queries channel closure. Since the channels are programmed to be closed 1,1
is returned as a string.
CLOS (@100,213) Close channels l00 and 2l3
CLOS? (@100,213) Query channels l00 and 2l3 state
OPEN [ROUTe:]OPEN <channel_list> opens the Form C Switch channels
specified by channel_list. channel_list has the form (@ccnn) where cc =
card number (01-99) and nn = channel number (00-15).
Parameters
Parameter Name Parameter Type Range of Values
<channel_list> numeric cc00 - cc15
Comments
Opening Channels: To open a single channel, use [ROUTe:]OPEN
(@ccnn); for multiple channels use [ROUTe:]OPEN (@ccnn,ccnn,);
for sequential channels, use [ROUTe:]OPEN (@ccnn:ccnn); for
groups of sequential channels, use [ROUTe:]OPEN
(@ccnn:ccnn,ccnn:ccnn); or any combination. Opening order for
multiple channels with a single command is not guaranteed.
Related Commands: [ROUTe:]CLOSe, [ROUTe:]OPEN?
*RST Condition: All channels open.
Example Opening Channels:
This example opens channels 100 and 213 of a two-module switchbox (card
numbers 01 and 02).
OPEN (@100,213) Open channels 100 and 213. 100
opens channel 00 of #1 and 213
opens channel 13 of card #2.
Chapter 5 Agilent E1364A Form C Switch Command Reference 53