Agilent Technologies E3632A Network Card User Manual


 
Chapter 4 Remote Interface Reference
Status Reporting Commands
110
*PSC?
This command queries the power-on status clear setting. The returned
parameter is ‘‘0’’
(*PSC 0
) or ‘‘1’’ (
*PSC 1
).
*SRE <enable value>
This command enables bits in the Status Byte enable register.
*SRE?
This command queries the Status Byte Enable register. The power supply
returns a decimal value which corresponds to the binary-weighted sum of all
bits set in the register.
*STB?
This command queries the Status Byte summary register. The
*STB?
command
is similar to a serial poll but it is processed like any other instrument command.
The
*STB?
command returns the same result as a serial poll but the ‘‘Request
Service’’ bit (bit 6) is not cleared if a serial poll has occurred.
*WAI
This command instructs the power supply to wait for all pending operations
to complete before executing any additional commands over the interface.
Used only in the triggered mode.