Agilent Technologies E3632A Network Card User Manual


 
Chapter 4 Remote Interface Reference
Calibration Commands
98
CALibration:VOLTage:LEVel {MINimum | MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output
state is ON. It sets the power supply to a calibration point that is entered with
CAL:VOLT
command. During calibration, three points must be entered and the
low-end point (MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of the power
supply. It takes about 7 seconds to execute the command. The calibration must
be unsecured and the output be opened before calibrating the overvoltage
protection circuit. The power supply automatically performs the calibration
and stores the new overvoltage constant in nonvolatile memory. Notice that
voltage calibration precedes before sending this command.