Chapter 5 Error Messages
Calibration Errors
129
5
Calibration Errors
The following errors indicate failures that may occur during a calibration.
Refer to the Service Guide for more information.
701
Cal security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
power supply. When applicable, this error will occur at power-on to warn you
that the power supply is unsecured.
702
Cal secured
The power supply is secured against calibration.
703
Invalid secure code
An invalid calibration security code was received when attempting to unsecure
or secure the power supply. You must use the same security code to unsecure
the power supply as was used to secure it, and vice versa. The security code
may contain up to 12 alphanumeric characters. The first character must be a
letter.
704
Secure code too long
A security code was received which contained more than 12 characters.
705
Cal aborted
A calibration in progress is aborted when you press any front-panel key, send
a device clear, or change the local/remote state of the instrument.
708
Cal output disabled
Calibration is aborted by sending
OUTP OFF
command during calibrating a
output.
712
Bad DAC cal data
The specified DAC calibration values (
CAL:VOLT
or
CAL:CURR
) are out of
range. Note that the new calibration constants are not stored in the non-volatile
memory.
713
Bad readback cal data
The specified readback calibration values (
CAL:VOLT
or
CAL:CURR
) are out
of range. Note that the new calibration constants are not stored in the non-
volatile memory.