Agilent Technologies E3632A Network Card User Manual


 
4
Front-Panel Voltage and Current Limit Settings
You can set the voltage and current limit values from the front panel using the
following method.
1 Select the desired range using the range selection keys after turning on the power
supply.
2 Press the key to show the limit values on the display.
3 Move the blinking digit to the appropriate position using the resolution selection keys
and change the blinking digit value to the desired voltage limit by turning the control
knob. If the display limit times out, press the key again.
4 Set the knob to current control mode using the voltage/current adjust selection key.
5 Move the blinking digit to the appropriate position using the resolution selection keys
and change the blinking digit value to the desired current limit by turning the control
knob.
6 Press the key to enable the output. After about 5 seconds, the
display will go to output monitoring mode automatically to display the voltage and
current at the output or the display will go to output monitoring mode immediately by
pressing the key again.
Note
All front panel keys and controls can be disabled with remote interface
commands. The Agilent E3632A must be in "Local" mode for the front panel
keys and controls to function.
Use the voltage/current adjust selection key, the resolution selection keys, and
the control knob to change the voltage and current limit values.
Display Limit
Display Limit
Output On/Off