HP (Hewlett-Packard) 2040D Network Router User Manual


 
Series 2040 Test System
Series 2040 Maintenance Manual V2.00Certification
46
Calibration Station. Finally, connect the GPIB port from the 2040 console to
the GPIB port on the Calibration Station.
To run the program select the
certification icon from the Atart/
Programs/Digalog program group,
and the screen to the right will
appear prompting for the type of
calibration. Select the TDAC, and
the screens in the right column on
both this page and the following
page will appear in the order
shown.
The first screen prompts for the
user’s name, the serial number of
the 8840 Fluke multimeter in the
Calibration Station. This number can be found on a Digalog tag inside the
Calibration Station next to the
meter. It also prompts for the
Selftest Assembly serial number,
and the serial number of the
tester itself.
When the information is
entered, the last setup screen
will appear requesting a
verification of the information
that was entered. If any of the
information is incorrect, it can
be modified or changed at this
time.
The operator’s name, Selftest
serial number, and Fluke 8840A
meter serial number must be
entered and verified before the
unit can be certified. Click on the “Certify” button and the program checks
the TDAC signal. Test results and a pass/fail condition are displayed in the