Bar Code Calibration Procedure without Remote Display*
10. Calibration process is now complete. The scanner will automatically restart, and beep one time, in
approximately 12 seconds.
11. Verify the calibration was successful with the calibrated weight. If the scale display does not accurately
show the calibrated load weight correctly, restart this procedure from Step 1 on page 53.
12. The unit’s calibration must now be verified as required by state and/or local Weight and Measures
regulations (starting on Page 58).
* These procedures are for MS2420/MS2430 Scanner/Diva Scale models only.