Calibrate Media and Ribbon Sensor Sensitivity
Use this procedure to adjust sensitivity of media and ribbon sensors.
Important • Follow this procedure exactly as presented. All of the
steps must be performed even if only one of the sensors requires
adjustment. You may press
MINUS (-) at any step in this procedure
to cancel the process.
To perform a media and ribbon sensor calibration:
1. Press SELECT to select the parameter.
2. Press PLUS (+) to start the calibration procedure.
The
LOAD BACKING prompt displays.
1. Open the printhead.
2. Remove approximately 8 in. (203 mm) of labels from the backing
(liner), and pull the media into the printer so that only the backing is
between the media sensors.
3. If your printer has the Peel-Off, Liner Take-Up, or Rewind option,
complete the following substeps:
a. Push down the peel-off mechanism release lever to open the peel
assembly.
b. Feed the liner over the tear-off/peel-off bar (1) and behind the peel
assembly (
2). Make sure that the end of the liner falls outside of
the printer.
Table 11 • Printer Parameters (Page 13 of 25)
Language/Parameter
Action/Explanation