Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 22
Example:
! Finish the Multi-Point Calibration function previously started (as indicated by the
previous examples of ‘C’ sub-commands ‘00’ and ‘01’). Calculate new adjustment
coefficients, and save them in the non-volatile memory of the module’s transducers.
These new coefficients will then be used for all subsequently calculated EU data
acquired by the module, until another calibration function is performed in the future.
“C 02”
“w08”
“w09”
Read responses (separately for each command executed above):
“A”
“A”
“A”
If an error (“N”) response is returned on the first command, either the correct
number of calibration points (per ‘00’ sub-command) were not supplied with
reasonable pressure data values (via the multiple ‘01’ sub-commands), or the
collected data yielded new calculated coefficients with unreasonable values. In that
case, the other two commands should not be used.
If execution of the last two ‘w’ commands is skipped above, the new calibration
data obtained will be stored only in volatile storage, and will be available for use
only until the module is RESET or loses power.