The coefficients of internal DH200 transducers used in the Model 9016 are selected
with array indexes aa=01 through 10 (hex). Coefficients of external 9400
transducers used by Models 9021 and 9022 (digitally compensated) are selected
with array indexes aa=01 through 0C (hex). All valid coefficient indexes (for each
of these arrays) are listed in the following table:
cc
Transducer Coefficient Description Datum Type
00
Re-zero Cal Adjustment (offset) term (Note 1)
FLOAT
01
Span Cal Adjustment (gain) term (Note 2)
FLOAT
02
Dynamic EU Conversion coefficient c0 (Note 3)
FLOAT
03
Dynamic EU Conversion coefficient c1 (Note 3)
FLOAT
04
Dynamic EU Conversion coefficient c2 (Note 3)
FLOAT
05
Dynamic EU Conversion coefficient c3 (Note 3)
FLOAT
07
User Defined Field (Note 4)
INTEGER
09
Transducer Manufacturing Reference Number (Note 5)
INTEGER
0A
Transducer Full-Scale Range Code (See Appendix F) (Note
5)
INTEGER
4D
Pressure Voltage Gain Index (Note 5)
INTEGER
4E Temperature Voltage Gain Index INTEGER
(Note 1) Related command ‘w08’ can be used to download the offset term to the sensor’s
non-volatile memory (digitally or non-digitally compensated sensor).
(Note 2) Related command ‘w09’ can be used to download the gain term to the sensor’s non-
volatile memory (digitally or non-digitally compensated sensor).
(Note 3) Related command ‘w27 can be used to download coefficients to the 9400 sensor’s
non-volatile memory (9021/9022 non-digitally compensated ONLY).
(Note 4) Data is immediately stored to the sensor’s non-volatile memory.
(Note 5) Data is immediately stored to the sensor’s non-volatile memory (9021/9022 non-