Pressure Systems, Inc. NetScanner™ System (9016, 9021, & 9022) User’s Manual
www.PressureSystems.com
3 - 40
Command ‘c’— Sub-command Index 03: Clear Stream(s)
This sub-command is used to “undefine” any previously configured host stream in a module, one
at a time, or all together. The sub-command’s format is:
Command “c 03 st”
‘c’ is the command letter
‘03’ is the sub-command index (‘ii’) for configuration
‘st’ is the stream identifier character (1, 2, or 3 or 0=all streams)
NOTE: all parameters are separated by a space character.
Response “A”
‘A’ is the acknowledge letter
Autonomous
Packet
none generated
Description: This sub-command clears (un-defines) the particular specified host stream (st=1-3),
or un-defines “all configured” host streams with a single command (st=0). Once
cleared, a stream must be reconfigured before it can be started.
Example:
! Stop all the streams configured previously. Then clear (un-define) only stream 3.
Finally, resume the remaining defined streams 1 and 2:
“c 02 0”
“c 03 3”
“c 01 0”
Read response:
“A”
“A”
“A”