Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration
This sub-command has no additional parameters.
Command “C 03”
‘C’ is the command letter.
‘ 03’ is the sub-command index (ii) for Abort.
NOTE: all parameters are separated by a space.
Response “A”
‘A’ is the acknowledge letter
Description: Aborts the Multi-Point Calibration function, if it is currently in progress. This
sub-command also restores the original “A/D samples for averaging” count to the
module that was in use before the calibration function was started.
It should be noted that executing the Configure & Start (‘00’) sub-command again,
after the calibration function has started collecting data (per Collect Data (‘01’) sub-
commands), but before the final data are calculated (per Calculate & Apply (‘02’)
sub-command), will have the same affect as this Abort function.
Example:
! Abort the Multi-Point Calibration function previously started
“C 03”
Read response:
“A”