Hach DR 5000 Scanner User Manual


 
114
Advanced Operations
3. Press Edit.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:
Stray Light
Photometrical accuracy
Wavelength accuracy
4. Press OK when all values are entered and the overview is
displayed.
5. Press Verification.
6. Insert the Multi Cell Holder in the cell compartment in such a
way that the cell opening for 10 mm square cuvettes/cells is at
the front. Secure the holder with two locking screws.
7. Remove any cuvettes/cells from the cell compartment, close
the cell compartment, and press Start.