114
Advanced Operations
3. Press Edit.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:
• Stray Light
• Photometrical accuracy
• Wavelength accuracy
4. Press OK when all values are entered and the overview is
displayed.
5. Press Verification.
6. Insert the Multi Cell Holder in the cell compartment in such a
way that the cell opening for 10 mm square cuvettes/cells is at
the front. Secure the holder with two locking screws.
7. Remove any cuvettes/cells from the cell compartment, close
the cell compartment, and press Start.