Standard Operations
5.5.6.2 Update an existing barcode test
Note: Only in very few cases does the revision of a test require all test
specifications to be updated.
Option 1: Manual update of test data
1. The instrument needs to be at the Main Menu.
2. Insert the blank or sample cuvette/vial (depending on the
working procedure) in cell compartment #1 and close the cell
compartment.
3. After "Barcode controlnumber?", a data overview is displayed,
including the specifications of the test that is to be revised.
Note: The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges, Conversion
Factor for chemical Form, etc.).
4. Highlight the corresponding line and press Edit.
5. When finished programming, press Store and then Cancel to
return to the User Programs menu.
6. Press Start to complete the test with the new data.
.
Option 2: Manual check/revision of test data
If an incorrect entry was made during the data update and is not
recognized via the barcode of the cuvette/sample vial (e.g. an
incorrect measuring range), another option is available to check
and if necessary to update, the test data.
1. The instrument need to be at the Main Menu.
2. Insert the blank or sample cuvette/vial (depending on the
working procedure) into the cell compartment #1 and close the
cell compartment.
3. Press Options>More>Edit.
4. A data overview is displayed, including the specifications of the
test that is to be revised. Compare the displayed data with the
data in the working procedure and edit the displayed data in
accordance with the working procedure.