93
Advanced Operations
6.5.2 Complete a measurement in the multi wavelength mode
1. Insert the blank cuvette/cell into the cuvette/sample cell holder.
Press Zero.
Note: The Read key does not become active until the zero
measurement has been completed.
2. Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press Read.
Note: The UV lamp is not switched on until a wavelength in the UV
spectrum has been entered and the blank reading has been started.
During the warm-up phase of the UV lamp, the message ”Warming
up...“ is displayed and the ”UV lamp“ symbol flashes. As soon as the
UV lamp is ready, the blank reading is carried out.
Note: Zero and Read are disabled until the cell compartment is
closed.
3. For data storage, see section 5.3.1.1 on page 35.