HP (Hewlett-Packard) HP 16500C Computer Accessories User Manual


 
“No Configuration File Loaded”
This is usually caused by trying to load a configuration file for one type of
module or the system into a different type of module.
Verify that the appropriate module has been selected from the Load
{module} from File {filename} in the HP 16500 disk operation menu.
Selecting Load {All} will cause incorrect operation when loading most
preprocessor interface configuration files.
See Also “To Load a Configuration File” in chapter 3, “File Management.”
“Selected File is Incompatible”
This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading the appropriate configuration file for your logic
analyzer.
“Slow or Missing Clock”
This error message might occur if the logic analyzer cards are not firmly
seated in the HP 16500C or HP 16501A frame. Ensure that the cards are
firmly seated.
This error might occur if the target system is not running properly.
Ensure that the target system is on and operating properly.
If the error message persists, check that the logic analyzer pods are
connected to the proper connectors. See the User’s Guide for your
preprocessor interface to determine the proper connections.
For HP 16510A Logic Analyzers, check the preprocessor interface power
fuse in the logic analyzer.
If You Have a Problem
“No Configuration File Loaded”
5–14