HP (Hewlett-Packard) HP 16500C Computer Accessories User Manual


 
To see the status of a module within an intermodule
measurement
1 Select the Intermodule menu.
2 Find the name of the module under the “Modules” list, and read the
status under the module name.
The status can be either Running or Stopped. You can interpret these
indications as follows:
If a module was running and is now stopped, assume it received its arming
signal, triggered, and finished its measurement properly.
If a module located below a stopped module on the intermodule
configuration tree has received an arming signal and is still running, either
it is still waiting to satisfy its trigger specification or it has not captured
enough information to fill its memory.
If a module below a running module on the intermodule configuration tree
has not received its arming signal, it will not begin running until the upper
module finds its trigger condition.
Intermodule Measurements
To see the status of a module within an intermodule measurement
2–10