HP (Hewlett-Packard) TIA/EIA-136 Network Card User Manual


 
65
Chapter 3, Processing Calls
Configure the Test Set to Emulate a Base Station
Processing Calls
CAUTION: SAVING
SETTINGS FOR PCS
OPERATION
To use SAVE/RECALL registers, you must use the same hardware configuration present
at the time the register is created. If you save instrument settings when a PCS Interface is
connected to the Test Set, the Test Set expects to detect the presence of the interface when
the register is recalled. If the PCS Interface is not connected to the Test Set when the reg-
ister is recalled, or the interface is turned off, a non-recoverable firmware error occurs. If
you encounter this error, the Test Set must be turned off, and the PCS Interface must be
connected and turned on, before the Test Set can power up normally and function correctly.
Emulating a Base
Station
A. How to access the DCCH CALL CONTROL screen
1. Verify that you have already done the following (refer to
Chapter 1, “Getting
Started,” on page 31
):
Zeroed the RF Power meter
Selected the proper audio filter, if you want to test audio
Connected the mobile to the Test Set
2. Select
CALL CNTL
from the
To Screen
. The CALL CONTROL screen will be
displayed.
3. Select
DCCH
from the
System Type
field. Now the CALL CONTROL screen shows
the parameters necessary to test TIA/EIA-136 mobiles.
B. Make the following settings on the CALL CONTROL screen:
System Type
:
DCCH
Define the settings for the control channel that you want your mobile to access on
power up:
Cntrl Chan
: enter your desired control channel number.
•Choose
Dig
if you want the DCCH
•Choose
Analog
if you want the ACC
•Choose
Cellular
to camp on a 850 MHz band control channel.
•Choose
PCS
if you want to camp on a 1900 MHz band control channel.
Amplitude
: enter the amplitude of the control channel.
SID
: enter the System Identification (in decimal format) of the base station to be emu-
lated.