Microtest TSB-67 - 568A Annex E Scanner User Manual


 
After calculating the NVP, press Save (F1) from the EDIT CABLE screen to store the new
settings. Then, press Default (F1) from the CABLE SETUP screen to select the cable as the
default cable type.
Customizing Autotest
To display the AUTOTEST SETUP screen, press the Extended Functions key, then select
Setup Autotest. This screen lets you set the default Autotest, edit Autotest settings and delete
user-defined Autotests.
Press the
Ñ or Ò arrow key to scroll through the list. Or, press the Ï or Ð arrow
key to scroll up or down one screen at a time. When the Autotest you want to use is highlighted,
press Default (F1) to select it as the Autotest to run.
If your Scanner has an alphanumeric overlay on the keypad, use the keypad to change the Autotest
name. If your Scanner does not have an alphanumeric overlay, press Edit (F2) to edit the name. See
Chapter 4: Editing, Printing & Updating for more information on editing. Press Change (F2)
to set the Margin Display.
Press Delete (F3) to delete user-defined Autotests.
Press the ESCAPE key to exit.
Editing Autotest Information
To display the EDIT AUTOTEST screen, press the Extended Functions key, then select Setup
Autotest, then highlight the Autotest to edit, then press Edit (F2). This screen lets you customize
Autotest settings.
Press the
Ñ or Ò arrow key to scroll through the available options and press Select (F1) to choose
the highlighted option. Or, press the number on the keypad that corresponds to the option you
want.
Select Name/Options to edit the Autotest name and to change the MARGIN OPTION. Select
Pairs to choose the pairs to be used in the Autotest, and to indicate which pairs have PASS/FAIL
criteria applied to them during Autotest. Select Tests to Run to choose which tests to run during
Autotest. Select Limits to edit test limits.
Press the ESCAPE key to exit.
Editing the Autotest Name
To display the EDIT AUTOTEST NAME screen, press the Extended Functions key, then select
Setup Autotests, then highlight the Autotest to edit, then press Edit (F2), then select
Name/Options. This screen lets you edit the Autotest name and change the MARGIN OPTION.
If your Scanner has an alphanumeric overlay on the keypad, use the keypad to enter up to 14
characters for the Autotest name. If your Scanner does not have an alphanumeric overlay, press
Edit (F2) to display the edit screen. See Chapter 4: Editing, Printing & Updating for more
information on editing.