National Instruments SCXI-1127 Switch User Manual


 
Appendix A Specifications for SCXI-1128
© National Instruments Corporation A-5 SCXI-1127/1128 User Manual
Dynamic Characteristics
Maximum operating speed.....................1200 cycles/s
Relay operate time (at 20 °C) ................ 0.25 ms typical, 0.5 ms max
Relay release time (at 20 °C)................. 0.08 ms typical, 0.2 ms max
Trigger Characteristics
Scanner advanced pulsewidth................1.1 µs
External trigger input pulsewidth........... 500 ns (min)
Stability
Recommended warm-up time................ 5 minutes
Physical
Dimensions.............................................3.0 by 17.2 by 20.3 cm
(1.2 by 6.9 by 8.0 in.)
Environment
Operating temperature..........................0 to 50 °C
Storage temperature..............................20 to 70 °C
Relative humidity...................................10% to 90% noncondensing
Electromagnetic compatibility (EMC)
emissions................................................ Refer to the document supplied
with the chassis for compliance to
relevant directives.
EMI ........................................................ FCC Class A compliant
Safety
Designed in accordance with IEC-61010-1, UL 3111-1, and
CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment
Approved at altitudes up to 2000 m.
Installation Category II