Chapter 3 SCT 320 Operation 35
2. On the Define Spectrograph dialog (Figure 3-6), click on the Install/Remove
Spectrograph button.
3. On the Install/Remove Spectrograph dialog, scroll down the list of supported
spectrographs, click on Acton SCT320, click on the Install Selected Spectrograph
button, and then click on OK to return to the Define Spectrograph dialog.
Figure 3-7. Install/Remove Spectrograph dialog
Figure 3-8. Define Spectrograph dialog: Acton SCT320 is Active Spectrograph
4. To confirm that WinSpec recognizes the spectrograph, click on Do Hardware Initialize.
You should hear the turret moving.
5. Now that you have identified the spectrograph to WinSpec, your next steps should be to
align the spectrograph optics to the detector. See “3.5.3 Perform Alignment” on the next
page.