2-12
SRX-T420
2-3. Optical Axis Adjustment of Lamp Bulb
Procedure
1. Be sure to click the “SERVICE” tab in the standby state.
The SERVICE window is displayed.
2. Click the [SELECT] button in a “LAMP ADJUST & TIMER RESET” menu.
LAMP INFORMATION window is displayed.
3. Enter the serial code described in the Operating Instructions of a lamp bulb in the “Serial Code” box
and click the [CHECK] button.
m
. Be sure to perform in the standby state.
. Enter the serial code with the space not put.
4. Click the [CLOSE] button and close the LAMP INFORMATION window.
5. Click the [ON] button of POWER.
“SERVICE” tab
[ON]
button
“LAMP ADJUST & TIMER RESET” menu
Fig. 2-3a
“LAMP ADJUST” menu
Enter the serial code.
[CHECK] button
[CLOSE] button
[START] button
Fig. 2-3b