Adjustment Procedures
5–12
TDS 684A, TDS 744A, & TDS 784A Service Manual
Figure 5–3: Performing Probe Compensation
2. Compensate the Probe:
a. Locate the probe compensation hole in the side of the probe body.
b. Using the probe compensation tool, adjust the probe for best square
wave compensation (flat as possible across its top).
Probe compensated correctly
Probe overcompensated
Probe undercompensated
Figure 5–4: Proper and Improper Probe Compensation