Agilent Technologies G6600-90006 Computer Accessories User Manual


 
Operation and Maintenance Manual 77
Instrument Shut-Down
Daily Shutdown
1 Toggle off the ozone generator.
2 Turn off the air regulator (counter-clockwise), located inside Detector door.
3 Toggle power to "stand-by."
4 Leave the vacuum pump and Dual Plasma Controller operating at all times.
Complete Shutdown
1 Toggle off the ozone generator.
2 Turn off the air regulator (counter-clockwise), located inside Detector door.
3 Toggle power to STANDBY.
4 Turn off power to the Dual Plasma Controller.
5 After 15 minutes, toggle off the vacuum pump, so the Burner cools and
moisture is removed from the system.
6 Turn off gases to unit.