44 Micro Motion EtherNet/IP Module
Device Profile
B.3.6 Output assembly for concentration measurement configu-
ration
B.3.7 Input parameters (explicit data) for concentration measure-
ment configuration
To update any of these attribute values, the associated trigger byte must be toggled before
reading the attribute value (executing the Get service). See Section B.3.4 for more information
on trigger bytes.
Byte Access Name Type Notes
0–3 Get/Set External Temperature Float
4–7 Get/Set External Pressure Float
8 Get/Set Start/Stop Totals Byte 0: Stop
1: Start
9 Get/Set Reset All Process Totals Byte 0: No action
1: Reset
10 Get/Set Reset All Inventory Totals Byte 0: No action
1: Reset
11 Get/Set Start Zero Byte 0: Abort or no action
1: Start
12–13 Get/Set Start Smart Meter Verifica-
tion
Word
See
Table B-20
Class Instance Attribute Access Name Type
Trigger
Byte Write
Attribute
Description
B0h 01h 01h Get Mass flow rate unit U16 0Ah
See
Table B-7
02h Get Density unit U16 0Ah
See
Table B-13
03h Get Temperature unit U16 0Ah
See
Table B-14
04h Get Volume flow rate unit U16 0Ah
See
Table B-9
05h Get Pressure unit U16 0Ah
See
Table B-15
06h Get Mass total/inventory unit U16 0Ah
See
Table B-8
07h Get Volume total/inventory unit U16 0Ah
See
Table B-10
08h Get Zero time U16 0Bh Seconds
09h Get Standard deviation of auto zero Float 0Ch
0Ah Get Present flow signal offset at
zero flow
Float 0Ch
0Bh Get Failed Zero Calibration Value Float 0Ch
0Ch Get Device Status Word 2 U16 0Dh See Section B.7.2