Emerson P/N MMI-20019808 Network Card User Manual


 
60 Micro Motion EtherNet/IP Module
Device Profile
Table B-22 Smart Meter Verification status
Bit number Status
0 State at abort
1
2
3
4Abort code
5
6
7Result:
0=Pass
1=Fail
Table B-23 Smart Meter Verification abort codes
Abort code Description Suggested action
1 User-initiated abort None required. Wait for 15 seconds before
starting another test.
3 Frequency drift Ensure that temperature, flow, and density
are stable, and rerun the test.
5 High drive gain Ensure that flow is stable, minimize
entrained gas, and rerun the test.
8 Unstable flow Enter that flow is stable and rerun the test.
13 No factory reference data for meter verification
test performed on air
Contact Micro Motion customer service
and provide the abort code.
14 No factory reference data for meter verification
test performed on water
Contact Micro Motion customer service
and provide the abort code.
15 No configuration data for meter verification Contact Micro Motion customer service
and provide the abort code.
Other General abort. Repeat the test. If the test aborts again,
contact Micro Motion customer service and
provide the abort code.
Table B-24 Smart Meter Verification state at abort
Value Description
0Inactive
1In progress
15 Debug mode
16 Measurement completed