HP (Hewlett-Packard) E1432A Graphics Tablet User Manual


 
Providing an External Clock
You can program the HP E1432A to accept an external word rate clock
from the Sample 0 line on the VXI Bus. The digital filters are still
functional, providing a range of effective word rates. All sampling is done
simultaneously and is not multiplexed.
To connect an External Sample Clock, use the External Sample SMB
connector on front panel of the HP E1432A. External Sample at word rate
and External Trigger are available on the front panel of HP E1432As which
do not have an arbitrary source or tachometer option.
The external clock must be a fixed frequency. Its maximum frequency
must not be higher than 100 kHz. Its minimum frequency must be at least
40.96 kHz.
HP E1432A User's Guide
Module Description
5-14