National Instruments AT E Series Network Card User Manual


 
Appendix A Specifications for AT-MIO-16XE-10 and AT-AI-16XE-10
© National Instruments Corporation A-23 AT E Series User Manual
CMRR, DC to 60 Hz
Dynamic Characteristics
Bandwidth
All gains.......................................... 255 kHz
Settling time for full-scale step
System noise (LSB
rms
including quantization noise)
Range CMRR (Bipolar) CMRR (Unipolar)
20 V 92 dB
10 V 97 dB 92 dB
5V 97 dB
4V 101 dB
2V 104 dB 101 dB
1V 105 dB 104 dB
100 mV to 500 mV 105 dB 105 dB
Accuracy*
±0.00076%
(±0.5 LSB)
±0.0015%
(±1 LSB)
±0.0016%
(±4 LSB)
40 µs max 20 µs max 10 µs max
* Accuracy values valid for source impedances < 1 k. Refer to the Multiple-Channel
Scanning Considerations section of Chapter 3, Hardware Overview,formore
information.
Range Bipolar Unipolar
2to20V 0.6 0.8
1V 0.7 0.8
400 to 500 mV 1.1 1.1
200 mV 2.0 2.0
100 mV 3.8