Appendix A Specifications for AT-MIO-16XE-50
AT E Series User Manual A-32 ni.com
CMRR, DC to 60 Hz
Dynamic Characteristics
Bandwidth
Settling time for full-scale step
Range CMRR (Bipolar) CMRR (Unipolar)
20 V 80 dB —
10 V 86 dB 80 dB
5V — 86 dB
2V 100 dB —
1V — 100 dB
200 mV 120 dB —
100 mV — 120 dB
Range Small Signal (–3dB)
5to20V 63 kHz
1to2V 57 kHz
100 to 200 mV 33 kHz
Range
Accuracy*
±0.0015%
(±1 LSB)
±0.0061%
(±4 LSB)
1to20V 50 µs max 50 µs max
200 mV (bipolar) 75 µs max 50 µs max
100 mV (unipolar) 75 µs max 50 µs max
* Accuracy values valid for source impedances < 1 kΩ. Refer to the Multiple-Channel
Scanning Considerations section of Chapter 3, HardwareOverview, for more information.