P6434 Probe Instructions
1
Operating Basics
The P6434 Mass Termination Probe allows you to connect 34 logic analyzer
channels to a microprocessor probe adapter or directly to your SUT in one fourth
the physical area needed to connect standard probes. The vertical clearance
dimension is less than a half inch when used with the Low-Profile Extender.
To connect directly to your SUT, you must include compatible Mictor connectors
in your circuit board design. To do this, be sure you understand the requirements
and conform to the guidelines for designing an interface in these instructions.
This section contains information on the following topics and tasks:
H Product description
H Applying labels
H Connecting probes to and disconnecting probes from a SUT
Product Description
The P6434 probe can acquire data from 32 input/output signals and two clock,
one clock and one qualifier, or two qualifier signals.
The probe has connectors on each end. One end is called the probe tip and
connects to your SUT. The other ends are called the module ends and connect to
the logic analyzer. Figure 1 shows the P6434 probe.
Probe tip
Module ends
Figure 1: P6434 Mass Termination Probe