Service Information
22
P6434 Probe Instructions
Disassembling the Probe
The following procedures describe disassembling the P6434 probe. If you are
installing the connector parts replacement kit, follow the Probe Tip procedures on
pages 22 and 25. If you are installing either of the cable replacement kits, you
must also perform the Latch Release Grip and Module End procedures.
To disassemble the probe tip, follow these steps (refer to Figure 15 as a guide):
Remove screws (4)
Elastomer connectors (2)
Circuit boards (2)
Figure 15: Disassembling the P6434 probe tip
1. Remove the four screws from the outside of the probe case using a .050 inch
hexagonal screwdriver, and open the case halves.
2. Gently remove the two small circuit boards and cables from the plastic
holder. The small circuit boards are held in place by two tab features in the
plastic holder.
The circuit boards hold the elastomer connectors (which are very small) in
place. When you remove the circuit boards, the elastomer connectors may
fall out of the probe.
Probe Tip