Table 4-1. Subsystem Commands Syntax (continued)
OUTPut
[:STATe] <bool> [,NORelay] Enables/disables the dc source output
:DFI
[:STATe] <bool> Enable/disable DFI output
:SOURce <source> Selects event source (QUES | OPER | ESB | RQS | OFF)
:PON
:STATe <state> Set power-on state (*RST | RCL0)
:PROTection
:CLEar Reset latched protection
:DELay <n> Delay after programming/before protection
:RELay
[:STATe] <bool> Opens/closes the external relay contacts
:POLarity <polarity> Sets the external relay polarity (NORM | REV)
:RI
:MODE <mode> Sets remote inhibit input (LATC | LIVE | OFF)
SENSe
:CURRent
[:DC]
RANGe [:UPPer] <n> Selects the high current measurement range
:DETector <detector> Selects the current measurement detector (ACDC | DC)
:FUNCtion <function> Configures the measurement sensor ("VOLT" | "CURR")
:SWEep
:OFFSet
:POINts <n> Defines the offset in the data sweep
:POINts <n> Define the number of data points in a sweep
:TINTerval <n> Sets the digitizer sample spacing
:WINDow [:TYPE] <type> Sets the measurement window function (HANN | RECT)
[SOURce:]
CURRent
[:LEVel]
[:IMMediate][:AMPLitude] <n> Sets the output current level
:TRIGgered [:AMPLitude] <n> Sets the triggered output current level
:PROTection
:STATe <bool> Enable/Disable current limit protection
DIGital
:DATA [:VALue] <n> Sets and reads the digital control port
:FUNCtion <function> Configures digital control port (RIDF | DIG)
VOLTage
[:LEVel]
[:IMMediate][:AMPLitude] <n> Sets the dc voltage level
:TRIGgered [:AMPLitude] <n> Sets the transient voltage level
:ALC
:BANDwidth? | :BWIDth? Returns setting of output mode switch
:PROTection [:LEVel] <n> Sets the overvoltage protection threshold