Agilent Technologies 66332A Power Supply User Manual


 
C - Error Messages
90
Table C-1. Error Numbers (continued)
Error
Number
Error String [Description/Explanation/Examples]
–141 Invalid character data [bad character, or unrecognized]
–144 Character data too long
–148 Character data not allowed
–150 String data error
–151 Invalid string data [e.g., END received before close quote]
–158 String data not allowed
–160 Block data error
–161 Invalid block data [e.g., END received before length satisfied]
–168 Block data not allowed
–170 Expression error
–171 Invalid expression
–178 Expression data not allowed
–200 Execution error [generic]
–222 Data out of range [e.g., too large for this device]
–223 Too much data [out of memory; block, string, or expression too long]
–224 Illegal parameter value [device-specific]
–225 Out of memory
–270 Macro error
–272 Macro execution error
–273 Illegal macro label
–276 Macro recursion error
–277 Macro redefinition not allowed
–310 System error
–350 Too many errors [errors beyond 9 lost due to queue overflow]
–400 Query error [generic]
–410 Query INTERRUPTED [query followed by DAB or GET before response complete]
–420 Query UNTERMINATED [addressed to talk, incomplete programming message received]
–430 Query DEADLOCKED [too many queries in command string]
–440 Query UNTERMINATED [after indefinite response]
0 No error
1 Non-volatile RAM RD0 section checksum failed
2 Non-volatile RAM CONFIG section checksum failed
3 Non-volatile RAM CAL section checksum failed
4 Non-volatile RAM STATE section checksum failed
5 Non-volatile RST section checksum failed
10 RAM selftest
11 VDAC/IDAC selftest 1
12 VDAC/IDAC selftest 2
13 VDAC/IDAC selftest 3
14 VDAC/IDAC selftest 4
15 OVDAC selftest
80 Digital I/O selftest error