Output Commands
Output commands consist of output and source commands.
Output commands control the output and digital port functions. They also control the output relay on
units with Relay Option 760.
Source commands program the actual voltage, current, and digital port output.
OUTPut
This command enables or disables the dc source output. The state of a disabled output is a condition of
zero output voltage and a model-dependent minimum source current (see *RST). Unless the NORelay
command is programmed, the OUTput command also controls the output relay on Agilent models
66332A, 6631B, 6632B, 6633B, and 6634B with Relay Option 760. If the NORelay command is sent, the
output relay state does NOT change.
Command Syntax
OUTPut[:STATe] <bool> [,NORelay]
Parameters
0 | OFF | 1 | ON
*RST Value
0
Examples
OUTP 1OUTPUT:STATE ON
Query Syntax
OUTPut[:STATe]?
Returned Parameters
<NR1>0 or 1
Related Commands
*RST *RCL *SAV
OUTPut:DFI
This command enables or disables the discrete fault indicator (DFI) outputfrom the dc source.
Command Syntax
OUTPut:DFI[:STATe]<bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
OUTP:DFI 1 OUTP:DFI ON
Query Syntax
OUTPut:DFI[:STATe]?
Returned Parameters
0 | 1
Related Commands
OUTP:DFI:SOUR
OUTPut:DFI:SOURce
This command selects the source for discrete fault indicator (DFI) events.The choices are:
QUEStionable
selects the Questionable event summary bit (bit 3 of the Status Byte Register)
OPERation
selects the Operation Event summary bit (bit 7 of the Status Byte Register)
ESB
selects the Standard Event summary bit (bit 5 of the Status Byte Register)
RQS
selects the Request Service bit (bit 6 of the Status Byte Register)
OFF
selects no DFI source
Command Syntax
OUTP:DFI:SOUR<source>
Parameters
QUES | OPER | ESB | RQS | OFF
*RST Value
OFF
Examples
OUTP:DFI:SOUR OPER
Query Syntax
OUTPut:DFI:SOUR?
Returned Parameters
<CRD>
Related Commands
OUTP:DFI