Agilent Technologies 66332A Power Supply User Manual


 
4 - Language Dictionary
46
CALibrate:STATe
This command enables and disables calibration mode. The calibration mode must be enabled before the
will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second parameter is the password. It is
required if the calibration mode is being enabled and the existing password is not 0. If the password is not
entered or is incorrect, an error is generated and the calibration mode remains disabled. The query
statement returns only the state, not the password.
NOTE: Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax
CALibrate:STATe<bool>[,<NRf>]
Parameters
0 | 1 | OFF | ON [,<password>]
*RST Value
OFF
Examples
CAL:STAT 1,6812 CAL:STAT OFF
Query Syntax
CALibrate:STATe?
Returned Parameters
<NR1>
Related Commands
CAL:PASS CAL:SAVE *RST
CALibrate:VOLTage
This command initiates the calibration of the output voltage and the voltage readback circuit.
Command Syntax
CALibrate:VOLTage[:DC]
Parameters
None
Examples
CAL:VOLT CAL:VOLT:DC
CALibrate:VOLTage:PROTection
This command can calibrates the overvoltage protection (OV) circuit. The dc source automatically
performs the calibration. CALibrate:VOLTage:PROTection is a sequential command that takes several
seconds to complete.
Command Syntax
CALibrate:VOLTage:PROTection
Parameters
None
Examples
CAL:VOLT:PROT