Agilent Technologies 66332A Power Supply User Manual


 
Language Dictionary - 4
77
TRIGger:SEQuence2:LEVel:CURRent
TRIGger:ACQuire:LEVel:CURRent
Agilent 66312A, 66332A Only
This command sets the trigger level for internally triggered current measurements. A positive current
trigger occurs when the current level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative current trigger occurs when the
current level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:LEVel:CURRent<NRf+>
TRIGger:ACQuire:LEVel:CURRent<NRf+>
Parameters
0 to MAX (see table 4-3)
Unit
A (amperes)
*RST Value
0
Examples
TRIG:SEQ2:LEV:CURR 5 TRIG:ACQ:LEV:CURR MAX
TRIG:ACQ:LEV 2
Query Syntax
TRIGger:SEQuence2:LEVel:CURRent?
TRIGger:ACQuire:LEVel:CURRent?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:LEV:VOLT TRIG:SEQ2:HYST:CURR
TRIGger:SEQuence2:LEVel:VOLTage
TRIGger:ACQuire:LEVel:VOLTage
Agilent 66312A, 66332A Only
This command sets the trigger level for internally triggered voltage measurements. A positive voltage
trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the
voltage level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:LEVel:VOLTage<NRf+>
TRIGger:ACQuire:LEVel:VOLTage<NRf+>
Parameters
0 to MAX (see table 4-3)
Unit
V (volts)
*RST Value
0
Examples
TRIG:SEQ2:LEV:VOLT 5 TRIG:ACQ:LEV:VOLT MAX
TRIG:ACQ:LEV 2
Query Syntax
TRIGger:SEQuence2:LEVel:VOLTage?
TRIGger:ACQuire:LEVel:VOLTage?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:LEV:CURR TRIG:SEQ2:HYST:VOLT