Agilent Technologies 66332A Power Supply User Manual


 
Language Dictionary - 4
53
SENSe:FUNCtion
Agilent 66312A, 66332A Only
This command configures the measurement sensor to measure either voltage or current when an acquire
trigger is used. The query returns the function setting, either VOLT or CURR.
Command Syntax
SENSe:FUNCtion <function
Parameters
"VOLTage" | "CURRent"
Examples
SENS:FUNC "VOLT"
Query Syntax
SENSe:FUNCtion?
Returned Parameters
<SRD>
SENSe:SWEep:OFFSet:POINts
Agilent 66312A, 66332A Only
This command defines the offset in a data sweep when an acquire trigger is used. Negative values
represent data samples taken prior to the trigger.
Command Syntax
SENSe:SWEep:OFFSet:POINts <NRf+>
Parameters
-4095 through 2,000,000,000
*RST Value
0
Examples
SENS:SWE:OFFS:POIN -2047
Query Syntax
SENSe:SWEep:OFFSet:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:POIN MEAS:ARR
SENSe:SWEep:POINts
This command defines the number of points in a data sweep.
Command Syntax
SENSe:SWEep:POINts<NRf+>
Parameters
0 through 4096
*RST Value
2048
Examples
SENS:SWE:POIN 1024
Query Syntax
SENSe:SWEep:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR
SENSe:SWEep:TINTerval
This command defines the time period between samples
Command Syntax
SENSe:SWEep:TINTerval<NRf+>
Parameters
15.6 microseconds through 31200 seconds
*RST Value
15.6 microseconds
Examples
SENS:SWE:TINT 31.2E-6
Query Syntax
SENSe:SWEep:TINTerval?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR