Quick Start: Learning How to Make Measurements
Learning to Make Transmission Measurements
Step 4. Output measurement results
This example procedure shows how to output (store) measurement results to a disk.
For more information on creating a hardcopy of the measurement results, refer to
the "Printing, Plotting, and Saving Measurement Results" chapter in the User's Guide.
CAUTION Do not mistake the line switch for the disk eject button. If the line switch is
mistakenly pushed, the instrument will be turned off, losing all settings and
data that have not been saved.
1. Insert a DOS- or LIF-formatted disk into the analyzer disk drive. The analyzer does not
support LIF-HFS (hierarchical file system).
2. Press . Choose to save the
measurement results to the analyzer's internal disk drive.
3. Press .
Toggle to ON if you want to store the error-corrected data on
disk with the instrument state.
Toggle to ON if you want to store the raw data (ratioed and
averaged, but no error-correction) on disk with the instrument state.
Toggle to ON if you want to store the formatted data on disk
with the instrument state.
Toggle to ON if you want to store user graphics on disk with
the instrument state.
Toggle to ON if you want to only store the measurement data
of the device under test. The analyzer will not store the instrument state and
error-correction (measurement calibration). Therefore, the saved data cannot be
retrieved into the analyzer.
NOTE Toggling to ON will override all of the other save
options. Because this type of data is only intended for computer
manipulation, the file contents of a save cannot be
recalled and displayed on the analyzer.
Choose if you want to store data in a binary format.
Choose if you want to store data in an ASCII format, to later
read on a computer.
4. Press and the analyzer saves the file with a default title.
Save/Recall
SELECT DISK
INTERNAL DISK
RETURN
DEFINE DISK-SAVE
DATA ARRAY on OFF
RAW ARRAY on OFF
FORMAT ARY on OFF
GRAPHICS on OFF
DATA ONLY on OFF
DATA ONLY on OFF
DATA ONLY ON off
SAVE USING BINARY
SAVE USING ASCII
RETURN