to be used to zero the analyzer, eliminating the need to switch external
tanks.
Overall Design.
The design of the sample system minimizes the volume of dead space, which
can retain residual gas from another route or previous mode of analysis.
Figure 2-2: Model 3160 flow schematic.
2.1.3 Electronics and Signal Processing
The Model 3160 analyzer has an embedded microcomputer, which controls all
signal processing, input/output, and display functions for the analyzer. System power
is supplied from a power supply module designed to be compatible with any interna-
tional power source.
The microcomputer, a liquid crystal display (LCD), and all analog signal pro-
cessing electronics are located inside a replaceable control module. A light emitting
diode (LED) display is located outside the control module.