to be used to zero the analyzer, eliminating the need to switch external
tanks.
• Overall Design.
The design of the sample system minimizes the volume of dead space, which
can retain residual gas from another route or previous mode of analysis.
Figure 2-2: Model 3160 flow schematic.
2.1.3 Electronics and Signal Processing
The Model 3160 analyzer has an embedded microcomputer, which controls all
signal processing, input/output, and display functions for the analyzer. System power
is supplied from a power supply module designed to be compatible with any interna-
tional power source.
The microcomputer, a liquid crystal display (LCD), and all analog signal pro-
cessing electronics are located inside a replaceable control module. A light emitting
diode (LED) display is located outside the control module.