Teledyne 3160 Computer Drive User Manual


 
4-5
Operations 4Operations 4
Operations 4Operations 4
Operations 4
TELEDYNE ANALYTICAL INSTRUMENTS
1. A cold start will automatically occur the first time you
start the analyzer. In any other situation, you can cold
start by turning the analyzer off and holding down the
red key below the LCD screen while you turn the power
back on. Continue to hold the red key until the screen
comes on. There will be a two minute wait for system
initialization before the second screen.
A warm start will take you directly to step 5.
2. Set the time using the SELEC, UP and DOWN keys,
and press ACCPT.
3. Select sensor A or B as the active sensor.
(This menu is omitted if only one cell block is installed.)
4. If you have not done so, connect the sample gas to the
sample port and press ACK.
5. Using the valve and meter on the sample gas supply, set
the flow rate to 2 SCFH and press ACK.
If you have warm-started, this is the second screen you
will see.
6. The main Analyze mode menu will appear, and the
measured O
2
level of the sample gas will display. The
system should first be zeroed, then spanned, to obtain
the most accurate O
2
measurements.
Cold Start-UpCold Start-Up
Cold Start-UpCold Start-Up
Cold Start-Up