Set-Up Alarms: Each of five alarms provided may be programmed to
actuate at any O
2
trace level above or below a threshold. The fifth alarm may
also be programmed as a system alarm to monitor several system-level condi-
tions. The system alarm can be triggered by the following system operations:
Scheduled zero (AutoZero) failure
Scheduled span (AutoSpan) failure
Battery failure
Internal errors
Any condition triggering the system alarm can be determined from status
page 4 (see this chapter: System Statistics).
When the alarms are disabled, scheduled zeroing or spanning will not
take place.
Logger Set-Up: A system data logger keeps a record of past oxygen
measurements. The logger is presented in the form of a graphed chart on the
system LCD; gaps in time will not display, and all points shown on a single
chart have the same range and sensor number.
The logger will not record in the Test Alternate Sensor mode.
Set Up Clock Functions: An internal clock provides the time and date
for events scheduled by the user. The internal clock is also used by the logger
to record events scheduled at timed intervals.
Scheduled zeroing and spanning (AutoZeroing and AutoSpanning) are set
in this mode. The mode uses an electronic timer to set off the zeroing or
spanning sequence at pre-selected intervals. During an AutoZero or AutoSpan,
the alarms are defeated and the screen displays PLEASE WAIT. SCHED-
ULED SPAN IN PROGRESS.
Since AutoZeroing and AutoSpanning uses automatic settling detection
of the sensor, which can take up to fifteen minutes, oxygen monitoring will be
interrupted for the length of time it takes to zero or span, after which monitor-
ing will resume. The LCD will display the current background level during
zeroing or spanning.
Changing Passwords For Remote Monitoring and Control:
Programmable security passwords are provided to prevent unauthorized access
to remote monitoring and control. Monitor is the lowest authorization level
and only permits monitoring of analyzer functions, while Control allows
remote control of any function.
System Statistics: Menu screens are available to view various system
parameters, and for setting serial link operation parameters, including modem
operation.