Agilent Technologies 1670G Printer Accessories User Manual


 
REMove
Command :MACHine{1|2}:TLISt:REMove
The REMove command removes all labels, except the leftmost label, from
the listing menu.
Example OUTPUT XXX;":MACHINE1:TLIST:REMOVE"
RUNTil (Run Until)
Command :MACHine{1|2}:TLISt:RUNTil <run_until_spec>
The RUNTil command defines a stop condition when the trace mode is
repetitive. Specifying OFF causes the analyzer to make runs until either
STOP is selected from the front panel or the STOP command is issued.
There are four conditions based on the time between the X and O markers:
The difference is less than (LT) some value
The difference is greater than (GT) some value
The difference is inside some range (INRange)
The difference is outside some range (OUTRange)
End points for the INRange and OUTRange should be at least 2 ns apart since
this is the minimum time between samples.
<run_until_
spec>
{OFF|LT,<value>|GT,<value>|INRange,<value>,<value>
|OUTRange,<value>,<value>}
<value>
real number from 9E9 to +9E9
Example
OUTPUT XXX;":MACHINE1:TLIST:RUNTIL GT,800.0E6"
Query :MACHine{1|2}:TLISt:RUNTil?
TLISt Subsystem
REMove
24–15